[Improvement of the surface coil of weida low magnetic field imaging system].

Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation · 1998 · Benefit · Insufficient evidence

← Research stream
BenefitInsufficient evidencePaper 4051

Replacing Litz wire with a copper tube (6 mm diameter, 0.7 mm thickness) increased the surface coil signal-to-noise ratio from 12 to 23 and the quality factor from 60 to 136, while reducing bandwidth from 50 kHz to 30 kHz in the WEIDA low magnetic field imagi…

Study snapshot

PublicationZhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation · 1998
Study typeengineering
Reported directionBenefit
Evidence ratingInsufficient

Abstract

By Replacing Litg wire with copper tube (phi 6 mm, thickness 0.7 mm), we increase the WEIDA low magnetic field imaging system surface coil's signal to noise ratio from 12 to 23, quality factor from 60 to 136, and reduce bandwidth from 50 kHz to 30 kHz.

Machine Enhanced Logic

AI evidence extraction

Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.

Study typeengineering
Effect directionbenefit
PopulationNot supplied
Sample sizeNot supplied
ExposureNot supplied
Evidence strengthInsufficientConfidence: 74% · Peer reviewed: yes

Main findings

Replacing Litz wire with a copper tube (6 mm diameter, 0.7 mm thickness) increased the surface coil signal-to-noise ratio from 12 to 23 and the quality factor from 60 to 136, while reducing bandwidth from 50 kHz to 30 kHz in the WEIDA low magnetic field imaging system.

Outcomes measured

  • signal-to-noise ratio
  • quality factor
  • bandwidth
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": null,
    "source": null,
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "signal-to-noise ratio",
    "quality factor",
    "bandwidth"
  ],
  "main_findings": "Replacing Litz wire with a copper tube (6 mm diameter, 0.7 mm thickness) increased the surface coil signal-to-noise ratio from 12 to 23 and the quality factor from 60 to 136, while reducing bandwidth from 50 kHz to 30 kHz in the WEIDA low magnetic field imaging system.",
  "effect_direction": "benefit",
  "limitations": [],
  "evidence_strength": "insufficient",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "surface coil",
    "low magnetic field imaging system",
    "coil design",
    "signal-to-noise ratio",
    "quality factor",
    "bandwidth",
    "copper tube",
    "Litz wire"
  ],
  "suggested_hubs": []
}
AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.

Identifiers and source

DOI: Not supplied
PubMed: 12019512

Read the original paper →

Related papers

← Back to the Research Library