Effects of Anatomical Differences on Electromagnetic Fields, SAR, and Temperature Change.
← Research streamNumerical simulations and a bovine-phantom experiment showed that transmit B-field distributions across different geometries/samples can be well correlated, while maximum SAR and temperature change (ΔT) can differ notably. The authors caution that relying on …
Study snapshot
Abstract
Electromagnetic field simulations are increasingly used to assure RF safety of patients during MRI exams. In practice, however, tissue property distribution of the patient being imaged is not known, but may be represented with a pre-existing model. Repeatedly, agreement in transmit magnetic (B) field distributions between two geometries has been used to suggest agreement in heating distributions. Here we examine relative effects of anatomical differences on B distribution, Specific Absorption Rate (SAR) and temperature change (ΔT). Numerical simulations were performed for a single surface coil positioned adjacent a homogeneous phantom and bovine phantom, each with slight geometric variations, and adjacent two different human body models. Experimental demonstration was performed on a bovine phantom using MR thermometry and B mapping. Simulations and experiments demonstrate that B distributions in different samples can be well correlated, while notable difference in maximum SAR and ΔT occur. This work illustrates challenges associated with utilizing simulations or experiments for RF safety assurance purposes. Reliance on B distributions alone for validation of simulations and/or experiments with a sample or subject for assurance of safety in another should be performed with caution.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
Numerical simulations and a bovine-phantom experiment showed that transmit B-field distributions across different geometries/samples can be well correlated, while maximum SAR and temperature change (ΔT) can differ notably. The authors caution that relying on B-field agreement alone to validate RF heating/safety across different subjects or models may be inadequate.
Outcomes measured
- Transmit magnetic (B) field distribution
- Specific Absorption Rate (SAR)
- Temperature change (ΔT)
Limitations
- Tissue property distribution of actual patients is not known and may be represented with pre-existing models
- Work focused on a single surface coil and specific phantoms/human body models; generalizability to other coils/geometries is not stated
- Experimental demonstration was performed on a bovine phantom (not in humans)
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": "RF",
"source": "MRI",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"Transmit magnetic (B) field distribution",
"Specific Absorption Rate (SAR)",
"Temperature change (ΔT)"
],
"main_findings": "Numerical simulations and a bovine-phantom experiment showed that transmit B-field distributions across different geometries/samples can be well correlated, while maximum SAR and temperature change (ΔT) can differ notably. The authors caution that relying on B-field agreement alone to validate RF heating/safety across different subjects or models may be inadequate.",
"effect_direction": "mixed",
"limitations": [
"Tissue property distribution of actual patients is not known and may be represented with pre-existing models",
"Work focused on a single surface coil and specific phantoms/human body models; generalizability to other coils/geometries is not stated",
"Experimental demonstration was performed on a bovine phantom (not in humans)"
],
"evidence_strength": "low",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"MRI",
"RF safety",
"electromagnetic field simulations",
"B-field mapping",
"MR thermometry",
"surface coil",
"phantom",
"anatomical differences",
"SAR",
"temperature change"
],
"suggested_hubs": [
{
"slug": "mri-rf-safety",
"weight": 0.9,
"reason": "Study evaluates RF fields, SAR, and heating (ΔT) in the context of MRI safety assurance using simulations and phantoms/human models."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1002/cmr.b.21317
PubMed: 27134586
Related papers
- Dependence of RF heating on SAR and implant position in a 1.5T MR system. 2007
- MRI induced heating of pacemaker leads: effect of temperature probe positioning and pacemaker placement on lead tip heating and local SAR. 2006
- Theoretical and experimental investigation of the relationship among SAR, tissues and radio frequencies in MRI. 2005
- Assessment of the specific absorption rate and calibration of decoupling parameters for proton decoupled carbon-13 MR spectroscopy at 3.0 T. 2005
- Thermal magnetic resonance: physics considerations and electromagnetic field simulations up to 23.5 Tesla (1GHz). 2015
- Wideband Self-Grounded Bow-Tie Antenna for Thermal MR. 2020