Absolute measurements of ultrasonic pressure by using high magnetic fields.
← Research streamA hydrophone concept is described that detects an EMF signal generated in a conductor when exposed to ultrasound in a uniform static magnetic field. Acoustic signals at 748 kHz were recorded in 1.5 T and 4.7 T fields, showing linear response over four orders …
Study snapshot
Abstract
A hydrophone is introduced that exploits the emf signal generated in a conductor when sonicated in the presence of a uniform static magnetic field. The method uses a small metal coil or metal membrane as a hydrophone receiver. Acoustic signals at 748 kHz are introduced in 1.5 T and 4.7 T fields and recorded both through direct electrical contact with the hydrophone and via RF pick-up coils, allowing wireless placement of the hydrophone. Linear response Is confirmed over four orders of magnitude in the pressure amplitude. Waveforms determined from the detected voltage are shown to be in excellent agreement with those obtained using a calibrated polyvinylidene difluoride film, and absolute values correlate within 20%. The methods are conceptually suitable for use in the presence of the high and uniform field of commercial MR scanners. The hydrophone may appear particularly useful as a quality assurance device in therapeutic and diagnostic acoustic techniques that use MRI.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
A hydrophone concept is described that detects an EMF signal generated in a conductor when exposed to ultrasound in a uniform static magnetic field. Acoustic signals at 748 kHz were recorded in 1.5 T and 4.7 T fields, showing linear response over four orders of magnitude in pressure amplitude, with waveforms in excellent agreement with a calibrated PVDF film and absolute values correlating within 20%.
Outcomes measured
- Hydrophone voltage response to acoustic pressure in static magnetic fields
- Linearity of pressure measurement
- Agreement with calibrated PVDF film hydrophone
- Feasibility of wireless placement via RF pick-up coils
Limitations
- No biological/health outcomes reported (engineering/measurement study).
- Sample size and experimental details beyond field strengths and frequency are not provided in the abstract.
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": null,
"source": "static magnetic field (MR-like) with ultrasound; RF pick-up coils used for signal detection",
"frequency_mhz": 0.748,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"Hydrophone voltage response to acoustic pressure in static magnetic fields",
"Linearity of pressure measurement",
"Agreement with calibrated PVDF film hydrophone",
"Feasibility of wireless placement via RF pick-up coils"
],
"main_findings": "A hydrophone concept is described that detects an EMF signal generated in a conductor when exposed to ultrasound in a uniform static magnetic field. Acoustic signals at 748 kHz were recorded in 1.5 T and 4.7 T fields, showing linear response over four orders of magnitude in pressure amplitude, with waveforms in excellent agreement with a calibrated PVDF film and absolute values correlating within 20%.",
"effect_direction": "unclear",
"limitations": [
"No biological/health outcomes reported (engineering/measurement study).",
"Sample size and experimental details beyond field strengths and frequency are not provided in the abstract."
],
"evidence_strength": "insufficient",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"hydrophone",
"ultrasonic pressure measurement",
"static magnetic field",
"1.5 T",
"4.7 T",
"MRI",
"electromotive force signal",
"RF pick-up coils",
"wireless measurement",
"quality assurance"
],
"suggested_hubs": [
{
"slug": "who-icnirp",
"weight": 0.2,
"reason": "Involves exposure to strong static magnetic fields comparable to MRI, which can relate to guideline discussions, though this is an engineering measurement paper."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1109/58.808874
PubMed: 18244347
Related papers
- Determination of blood longitudinal relaxation time (T1) at high magnetic field strengths. 2007
- Electrophysiological modeling study of ECG T-wave alternation caused by ultrahigh static magnetic fields. 2005
- Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head. 2002
- [A study on the effects of RF coil for the static magnetic field]. 2001
- Effects of a 4.7 T static magnetic field on fetal development in ICR mice. 2001
- Static magnetic fields generated by a 0.5 T MRI unit affects in vitro expression of activation markers and interleukin release in human peripheral blood mononuclear cells (PBMC). 1999