Radiofrequency Interference in the Clinical Laboratory
← Research streamAn increase in power output from an antenna located ~4 feet from a hematology analyzer was temporally associated with aberrant side scatter and abnormal/erroneous WBC counts. Powering down the antenna returned the instrument to normal working conditions, and …
Study snapshot
Abstract
Radiofrequency Interference in the Clinical Laboratory Badizadegan ND, Greenberg S, Lawrence H, Badizadegan K. Radiofrequency Interference in the Clinical Laboratory. Am J Clin Pathol. 2019 Apr 2;151(5):522-528. doi: 10.1093/ajcp/aqy174. Abstract OBJECTIVES: Radiofrequency interference (RFI) is a known medical device safety issue, but there are no documented cases of interference resulting in erroneous laboratory results. METHODS: We investigated unexpected failure of a hematology analyzer resulting in erroneous WBC counts. Hardware failure was initially suspected, but temporal association with increased power output from a nearby antenna prompted investigation for RFI. RESULTS: Power output from an antenna located approximately 4 feet from the analyzer was increased to ensure sufficient signal for emergency communications in the building. Interference from the antenna resulted in aberrant side scatter and abnormal WBC counts. Powering down the antenna returned the instrument to normal working conditions. CONCLUSIONS: We have shown RFI as the root cause of erroneous WBC counts in a hematology analyzer. We propose that RFI should be on the list of potential interfering mechanisms when clinical laboratory instruments generate inconsistent or unreliable results. Open access paper: ncbi.nlm.nih.gov
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
An increase in power output from an antenna located ~4 feet from a hematology analyzer was temporally associated with aberrant side scatter and abnormal/erroneous WBC counts. Powering down the antenna returned the instrument to normal working conditions, and the authors conclude radiofrequency interference was the root cause.
Outcomes measured
- hematology analyzer performance
- WBC (white blood cell) count accuracy
- side scatter signal abnormalities
Limitations
- Single instrument/event; generalizability unclear
- No RF frequency or quantitative exposure measurements reported in abstract
- Potential alternative causes not fully described in abstract
View raw extracted JSON
{
"study_type": "case_report",
"exposure": {
"band": "RF",
"source": "nearby antenna (emergency communications)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"hematology analyzer performance",
"WBC (white blood cell) count accuracy",
"side scatter signal abnormalities"
],
"main_findings": "An increase in power output from an antenna located ~4 feet from a hematology analyzer was temporally associated with aberrant side scatter and abnormal/erroneous WBC counts. Powering down the antenna returned the instrument to normal working conditions, and the authors conclude radiofrequency interference was the root cause.",
"effect_direction": "harm",
"limitations": [
"Single instrument/event; generalizability unclear",
"No RF frequency or quantitative exposure measurements reported in abstract",
"Potential alternative causes not fully described in abstract"
],
"evidence_strength": "very_low",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"radiofrequency interference",
"RFI",
"clinical laboratory",
"hematology analyzer",
"white blood cell count",
"medical device safety",
"antenna",
"emergency communications"
],
"suggested_hubs": [
{
"slug": "medical-device-interference",
"weight": 0.9,
"reason": "Reports radiofrequency interference causing erroneous results in a hematology analyzer due to a nearby antenna."
},
{
"slug": "occupational-exposure",
"weight": 0.25,
"reason": "Exposure source is an antenna in a building setting affecting laboratory equipment; relevance is indirect (device interference rather than health outcome)."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1093/ajcp/aqy174
PubMed: Not supplied
Related papers
- Time-dependent hematological changes in workers exposed to electromagnetic fields. 1995
- Radiation Effect Analysis on Human Head Caused by Different Frequencies of Mobile Phone Antenna 2022
- A novel mobile phone antenna for effectively reducing specific absorption rate 2018
- Mortality by neoplasia and cellular telephone base stations in the Belo Horizonte municipality, Minas Gerais state, Brazil. 2011
- Investigation of the spinal cord as a natural receptor antenna for incident electromagnetic waves and possible impact on the central nervous system. 2012
- Electromagnetic field pattern in the environment of GSM base stations. 1999