The effect of radar on cardiac pacemakers.
← Research streamIn laboratory tests in air, all 16 pacemakers showed signs of interference at pulse power densities between 0.025 mW/cm2 and 62.5 mW/cm2. In field testing 1.2 km from the radar station, 3 of 6 implanted pacemakers were triggered or inhibited depending on mode…
Study snapshot
Abstract
The susceptibility of 16 noncompetitive cardiac pacemakers to radiation from a powerful radar system was investigated in the laboratory and in the vicinity of its prototype. From comparative in vitro tests in air, fat, water, and saline it was concluded that only tests in fat or air represent the worst case condition after implantation. In air all pacemakers showed signs of interference at pulse power densities between 0.025 mW/cm2 and 62.5 mW/CM2. Three of six implanted pacemakers were triggered or inhibited depending on their mode of operation when tested at a location 1.2 km away from the radar station by the radar beam occurring every 5.5 sec. Because interfering radiation can enter the pacemaker circuitry directly along the electrode, acting as an antenna, metal encapsulation of the pulse generator does not provide sufficient shielding against microwave radiation. However, pacemakers modified by metal encapsulation and a low-pass filter at the electrode remained undistrubed at pulse power densities of greater than 10 W/cm2 when tested under worst case condition in air.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
In laboratory tests in air, all 16 pacemakers showed signs of interference at pulse power densities between 0.025 mW/cm2 and 62.5 mW/cm2. In field testing 1.2 km from the radar station, 3 of 6 implanted pacemakers were triggered or inhibited depending on mode of operation. Pacemakers modified with metal encapsulation plus a low-pass filter at the electrode reportedly remained undisturbed at pulse power densities greater than 10 W/cm2 under worst-case conditions in air.
Outcomes measured
- pacemaker electromagnetic interference (triggering/inhibition)
- susceptibility under different test media (air, fat, water, saline)
- effect of shielding/modifications (metal encapsulation, low-pass filter)
Limitations
- Frequency of the radar signal not reported in the abstract.
- Exposure characterization is primarily in terms of pulse power density; other metrics (e.g., SAR) not provided.
- Small number of implanted pacemakers tested in the field (6).
- Laboratory conditions (air/fat/water/saline) may not fully represent in vivo implantation conditions.
View raw extracted JSON
{
"study_type": "other",
"exposure": {
"band": "microwave",
"source": "radar",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": 16,
"outcomes": [
"pacemaker electromagnetic interference (triggering/inhibition)",
"susceptibility under different test media (air, fat, water, saline)",
"effect of shielding/modifications (metal encapsulation, low-pass filter)"
],
"main_findings": "In laboratory tests in air, all 16 pacemakers showed signs of interference at pulse power densities between 0.025 mW/cm2 and 62.5 mW/cm2. In field testing 1.2 km from the radar station, 3 of 6 implanted pacemakers were triggered or inhibited depending on mode of operation. Pacemakers modified with metal encapsulation plus a low-pass filter at the electrode reportedly remained undisturbed at pulse power densities greater than 10 W/cm2 under worst-case conditions in air.",
"effect_direction": "harm",
"limitations": [
"Frequency of the radar signal not reported in the abstract.",
"Exposure characterization is primarily in terms of pulse power density; other metrics (e.g., SAR) not provided.",
"Small number of implanted pacemakers tested in the field (6).",
"Laboratory conditions (air/fat/water/saline) may not fully represent in vivo implantation conditions."
],
"evidence_strength": "low",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"radar",
"microwave radiation",
"cardiac pacemaker",
"electromagnetic interference",
"pulse power density",
"shielding",
"low-pass filter",
"implantable medical device"
],
"suggested_hubs": [
{
"slug": "occupational-exposure",
"weight": 0.35,
"reason": "Radar exposure context may relate to occupational/operational environments where radar systems are present."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: Not supplied
PubMed: 1176271
Related papers
- Dosimetry of electromagnetic field exposure of an active armlet and its electromagnetic interference to the cardiac pacemakers using adult, child and infant models. 2016
- Interference with Implanted Upper Airway Stimulation Device by Phones with Magnet Technology (iPhones 12 and 13) 2022
- Electromagnetic Interference in the Modern Era: Concerns, Trends, and Nanomaterial-Based Solutions 2025
- Environmental Factors as the Main Hormonal Disruptors of Male Fertility 2024
- Occupational Exposure to Nonionizing Radiation and Risk for Malignancy in Young Adults 2023
- Human Health Risk Assessment of 4-12 GHz Radar Waves using CST STUDIO SUITE Software 2022