In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators.

Heart rhythm · 2010 · Harm · Low evidence

← Research stream
HarmLow evidencePaper 3367

In vitro, reactions consistent with EMI were observed during exposure to low-frequency (134 kHz) RFID in 67% of pacemaker tests (maximum distance 60 cm) and 47% of ICD tests (maximum distance 40 cm). During high-frequency (13.56 MHz) RFID exposure, reactions …

Study snapshot

PublicationHeart rhythm · 2010
Study typein_vitro
Reported directionHarm
Evidence ratingLow

Abstract

BACKGROUND: The use of radiofrequency identification (RFID) systems is expanding and highlights the need to address electromagnetic interference (EMI) to implantable pacemakers and implantable cardioverter-defibrillators (ICDs). OBJECTIVE: This study sought to examine the electromagnetic compatibility (EMC) between RFID readers and implantable pacemakers or ICDs. METHODS: During in vitro testing, 15 implantable pacemakers and 15 ICDs were exposed to 13 passive RFID readers in 3 frequency bands: 134 kHz (low frequency [LF]), 13.56 MHz (high frequency [HF]), and 915 MHz (ultra high frequency [UHF]). RESULTS: While being exposed to LF RFID, a reaction was observed for 67% of all pacemaker tests (maximum distance 60 cm) and 47% of all ICD tests (maximum distance 40 cm). During HF RFID exposure, a reaction was observed for 6% of all pacemaker tests (maximum distance 22.5 cm) and 1% of all ICD tests (maximum distance 7.5 cm). For both pacemakers and ICDs, no reactions were observed during exposure to UHF RFID or continuous-wave RFID. Pacemakers and ICDs were most susceptible to modulated LF RFID readers. CONCLUSION: Although there is in vitro testing evidence for concern for implantable pacemaker and ICD EMI at LF and HF, the FDA has not received any incident reports of pacemaker or ICD EMI caused by any RFID system. We do not believe the current situation reveals an urgent public health risk.

Machine Enhanced Logic

AI evidence extraction

Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.

Study typein_vitro
Effect directionharm
PopulationImplantable pacemakers and implantable cardioverter-defibrillators (ICDs) tested in vitro
Sample size30
ExposureRF
Evidence strengthLowConfidence: 78% · Peer reviewed: yes

Main findings

In vitro, reactions consistent with EMI were observed during exposure to low-frequency (134 kHz) RFID in 67% of pacemaker tests (maximum distance 60 cm) and 47% of ICD tests (maximum distance 40 cm). During high-frequency (13.56 MHz) RFID exposure, reactions were observed in 6% of pacemaker tests (maximum distance 22.5 cm) and 1% of ICD tests (maximum distance 7.5 cm). No reactions were observed during exposure to ultra-high-frequency (915 MHz) RFID or continuous-wave RFID; devices were most susceptible to modulated LF RFID readers.

Outcomes measured

  • Electromagnetic interference (EMI) reactions in pacemakers
  • Electromagnetic interference (EMI) reactions in ICDs
  • Maximum distance at which reactions observed

Limitations

  • In vitro testing (not in vivo/clinical setting)
  • Exposure details beyond frequency bands (e.g., field strength) not provided in abstract
  • Outcome described as 'reaction' without further definition in abstract
  • Authors note lack of FDA incident reports and do not consider it an urgent public health risk
View raw extracted JSON
{
  "study_type": "in_vitro",
  "exposure": {
    "band": "RF",
    "source": "RFID reader",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": "Implantable pacemakers and implantable cardioverter-defibrillators (ICDs) tested in vitro",
  "sample_size": 30,
  "outcomes": [
    "Electromagnetic interference (EMI) reactions in pacemakers",
    "Electromagnetic interference (EMI) reactions in ICDs",
    "Maximum distance at which reactions observed"
  ],
  "main_findings": "In vitro, reactions consistent with EMI were observed during exposure to low-frequency (134 kHz) RFID in 67% of pacemaker tests (maximum distance 60 cm) and 47% of ICD tests (maximum distance 40 cm). During high-frequency (13.56 MHz) RFID exposure, reactions were observed in 6% of pacemaker tests (maximum distance 22.5 cm) and 1% of ICD tests (maximum distance 7.5 cm). No reactions were observed during exposure to ultra-high-frequency (915 MHz) RFID or continuous-wave RFID; devices were most susceptible to modulated LF RFID readers.",
  "effect_direction": "harm",
  "limitations": [
    "In vitro testing (not in vivo/clinical setting)",
    "Exposure details beyond frequency bands (e.g., field strength) not provided in abstract",
    "Outcome described as 'reaction' without further definition in abstract",
    "Authors note lack of FDA incident reports and do not consider it an urgent public health risk"
  ],
  "evidence_strength": "low",
  "confidence": 0.78,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "RFID",
    "electromagnetic interference",
    "EMI",
    "electromagnetic compatibility",
    "EMC",
    "pacemaker",
    "implantable cardioverter-defibrillator",
    "ICD",
    "low frequency",
    "high frequency",
    "ultra high frequency"
  ],
  "suggested_hubs": [
    {
      "slug": "medical-implants-emi",
      "weight": 0.95,
      "reason": "Study tests EMI from RFID readers on implantable pacemakers and ICDs."
    },
    {
      "slug": "rfid",
      "weight": 0.85,
      "reason": "Exposure source is passive RFID readers across LF/HF/UHF bands."
    }
  ]
}
AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.

Identifiers and source

DOI: 10.1016/j.hrthm.2009.09.071
PubMed: 20129290

Read the original paper →