Radio frequency induced heating of an insulated wire during magnetic resonance imaging.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2016 · Unclear · Low evidence

← Research stream
UnclearLow evidencePaper 2882

3-D electromagnetic simulations assessed RF-induced power deposition near straight insulated leads under different incident electric field profiles. For the investigated setups, up to a 400% discrepancy was observed between implant model predictions and EM si…

Study snapshot

PublicationAnnual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2016
Study typeengineering
Reported directionUnclear
Evidence ratingLow

Abstract

Radio frequency (RF) induced power deposition near straight insulated leads excited by different incident electrical field profiles was investigated. 3-D electromagnetic (EM) simulations were used to obtain implant models as well as RF-induced power depositions calculated as the integral of power loss density. For the investigated setups up to 400% discrepancy between the implant model predictions and EM simulation values was observed.

Machine Enhanced Logic

AI evidence extraction

Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.

Study typeengineering
Effect directionunclear
PopulationNot supplied
Sample sizeNot supplied
ExposureRF
Evidence strengthLowConfidence: 74% · Peer reviewed: yes

Main findings

3-D electromagnetic simulations assessed RF-induced power deposition near straight insulated leads under different incident electric field profiles. For the investigated setups, up to a 400% discrepancy was observed between implant model predictions and EM simulation values.

Outcomes measured

  • RF-induced power deposition near insulated leads
  • Discrepancy between implant model predictions and 3-D EM simulation values

Limitations

  • No MRI/RF frequency reported in the abstract
  • No experimental (in vivo/in vitro) validation described; simulations only
  • No quantitative heating/temperature outcomes reported (power deposition only)
  • Sample size and specific lead/implant configurations not detailed
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": "RF",
    "source": "magnetic resonance imaging (MRI)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": null
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "RF-induced power deposition near insulated leads",
    "Discrepancy between implant model predictions and 3-D EM simulation values"
  ],
  "main_findings": "3-D electromagnetic simulations assessed RF-induced power deposition near straight insulated leads under different incident electric field profiles. For the investigated setups, up to a 400% discrepancy was observed between implant model predictions and EM simulation values.",
  "effect_direction": "unclear",
  "limitations": [
    "No MRI/RF frequency reported in the abstract",
    "No experimental (in vivo/in vitro) validation described; simulations only",
    "No quantitative heating/temperature outcomes reported (power deposition only)",
    "Sample size and specific lead/implant configurations not detailed"
  ],
  "evidence_strength": "low",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "radiofrequency",
    "MRI",
    "RF-induced heating",
    "power deposition",
    "insulated wire",
    "leads",
    "electromagnetic simulation",
    "incident electric field"
  ],
  "suggested_hubs": [
    {
      "slug": "occupational-exposure",
      "weight": 0.2,
      "reason": "MRI-related RF exposure context, but no specific occupational or human exposure details provided."
    }
  ]
}
AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.

Identifiers and source

DOI: 10.1109/embc.2016.7592154
PubMed: 28269677

Read the original paper →