New-generation electronic appliances and cardiac implantable electronic devices: a systematic
← Research streamTen in vivo studies were included (EVs, smartphones, smartwatches). No EMI incidence was reported with EVs or smartwatches. Magnet-containing smartphones (iPhone 12) could cause EMI when placed directly over CIEDs, triggering magnet mode; otherwise no EMI was…
Study snapshot
Abstract
New-generation electronic appliances and cardiac implantable electronic devices: a systematic literature review of mechanisms and in vivo studies Kewcharoen J, Shah K, Bhardwaj R, Contractor T, Turagam MK, Mandapati R, Lakkireddy D, Garg J. New- generation electronic appliances and cardiac implantable electronic devices: a systematic literature review of mechanisms and in vivo studies. J Interv Card Electrophysiol. 2024 Mar 5. doi: 10.1007/s10840-024-01777-z. Abstract Introduction: Cardiac implantable electronic device (CIED) functions are susceptible to electromagnetic interference (EMI) from electromagnetic fields (EMF). Data on EMI risks from new-generation electronic appliances (EA) are limited. Objective: We performed a systematic literature review on the mechanisms of EMI, current evidence, and recently published trials evaluating the effect of EMF on CIEDs from electric vehicles (EV), smartphone, and smartwatch technology and summarize its safety data. Methods: Electronic databases, including PubMed and EMBASE, were searched for in vivo studies evaluating EMF strength and incidence between CIEDs and commercial EVs, new-generation smartphones, and new-generation smartwatches. Results: A total of ten studies (three on EVs, five on smartphones, one on smartphones, one on smartphones and smartwatches) were included in our systematic review. There was no report of EMI incidence associated with EVs or smartwatches. Magnet-containing smartphones (iPhone 12) can cause EMI when placed directly over CIEDs - thereby triggering the magnet mode; otherwise, no report of EMI was observed with other positions or smartphone models. Conclusion: Current evidence suggests CIED recipients are safe from general interaction with EVs/HEVs, smartphones, and smartwatches. Strictly, results may only be applied to commercial brands or models tested in the published studies. There is limited data on EMI risk from EVs wireless charging and smartphones with MagSafe technology. pubmed.ncbi.nlm.nih.gov
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
Ten in vivo studies were included (EVs, smartphones, smartwatches). No EMI incidence was reported with EVs or smartwatches. Magnet-containing smartphones (iPhone 12) could cause EMI when placed directly over CIEDs, triggering magnet mode; otherwise no EMI was observed with other positions or smartphone models.
Outcomes measured
- Electromagnetic interference (EMI) incidence with CIEDs
- EMF strength in relation to CIEDs
- Triggering of magnet mode in CIEDs
Limitations
- Limited data on EMI risks from new-generation electronic appliances
- Findings may only apply to the specific commercial brands/models tested in included studies
- Limited data on EMI risk from EV wireless charging
- Limited data on EMI risk from smartphones with MagSafe technology
View raw extracted JSON
{
"study_type": "systematic_review",
"exposure": {
"band": null,
"source": "electric vehicles; smartphones; smartwatches",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": "CIED recipients (cardiac implantable electronic device users)",
"sample_size": null,
"outcomes": [
"Electromagnetic interference (EMI) incidence with CIEDs",
"EMF strength in relation to CIEDs",
"Triggering of magnet mode in CIEDs"
],
"main_findings": "Ten in vivo studies were included (EVs, smartphones, smartwatches). No EMI incidence was reported with EVs or smartwatches. Magnet-containing smartphones (iPhone 12) could cause EMI when placed directly over CIEDs, triggering magnet mode; otherwise no EMI was observed with other positions or smartphone models.",
"effect_direction": "mixed",
"limitations": [
"Limited data on EMI risks from new-generation electronic appliances",
"Findings may only apply to the specific commercial brands/models tested in included studies",
"Limited data on EMI risk from EV wireless charging",
"Limited data on EMI risk from smartphones with MagSafe technology"
],
"evidence_strength": "low",
"confidence": 0.78,
"peer_reviewed_likely": "yes",
"keywords": [
"cardiac implantable electronic devices",
"CIED",
"electromagnetic interference",
"EMI",
"electromagnetic fields",
"EMF",
"electric vehicles",
"EV",
"hybrid electric vehicles",
"smartphone",
"smartwatch",
"MagSafe",
"iPhone 12",
"magnet mode"
],
"suggested_hubs": [
{
"slug": "smart-meters",
"weight": 0,
"reason": null
},
{
"slug": "occupational-exposure",
"weight": 0,
"reason": null
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1007/s10840-024-01777-z
PubMed: 38443707
Related papers
- Effects of medically generated electromagnetic interference from medical devices on cardiac implantable electronic devices: A review 2022
- Security millimetre wave body scanner safe for patients with leadless pacemakers or subcutaneous implantable cardioverter-defibrillators 2020
- Patients with pacemakers or defibrillators do not need to worry about e-Cars: An observational study 2019
- Safety of smartwatches and their chargers in patients with cardiac implantable electronic devices 2020
- Impact of magnetic fields from tablets, laptops, smartphones, and household/leisure magnets on cardiac implantable electronic devices 2025