How much are anesthesiologists exposed to electromagnetic fields in operating rooms?
← Research streamMagnetic field density measured in 19 operating rooms averaged 2.22 +/- 1.13 mG at 30 cm from the main monitor center, 1.29 +/- 0.84 mG at 50 cm, and 1.00 +/- 0.78 mG at the anesthesiologist's standing point. Using 2 mG as a cutoff for high exposure, some ope…
Study snapshot
Abstract
Numerous electronic devices have been introduced into the operating room. Although little is known about the relationship between exposure to electromagnetic fields (EMF) and health hazards, some authors reported its association with cancer or other diseases. We measured the amount of EMF exposure that an anesthesiologist gets in the operating room. The density of the magnetic field was checked by an extremely low frequency (ELF) field strength measurement system in the 19 operating rooms of our hospital. We measured the magnetic field intensity at a distance of 30 cm, 50 cm, and at the place where the anesthesiologist usually stands from the center of the main monitor. The average exposure quantities of magnetic fields in 19 operating rooms were 2.22 +/- 1.13 mG at 30 cm, 1.29 +/- 0.84 mG at 50 cm and 1.00 +/- 0.78 mG at the anesthesiologist's standing points respectively. Because quantities over 2 or 3 mG were accepted to be high radiation levels of EMF by many reports describing the hazards of EMF, we set 2 mG to be the cutoff value. In some of the 19 operating rooms, the measured EMF density exceeded our cutoff value. Although the health hazards related to EMF exposure are still equivocal, anesthesiologists should consider making an effort to improve their environment and reduce their exposure to EMF.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
Magnetic field density measured in 19 operating rooms averaged 2.22 +/- 1.13 mG at 30 cm from the main monitor center, 1.29 +/- 0.84 mG at 50 cm, and 1.00 +/- 0.78 mG at the anesthesiologist's standing point. Using 2 mG as a cutoff for high exposure, some operating rooms exceeded this level.
Outcomes measured
- Magnetic field density/intensity (mG) at specified distances/locations in operating rooms
Limitations
- Health outcomes were not measured (exposure assessment only).
- Cutoff value (2 mG) was based on prior reports and not justified within the study.
- Frequency characteristics beyond 'ELF' were not reported.
- Single-hospital setting may limit generalizability.
View raw extracted JSON
{
"study_type": "exposure_assessment",
"exposure": {
"band": "ELF",
"source": "occupational (operating room electronic devices/monitor)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": "Anesthesiologists in operating rooms (hospital setting)",
"sample_size": 19,
"outcomes": [
"Magnetic field density/intensity (mG) at specified distances/locations in operating rooms"
],
"main_findings": "Magnetic field density measured in 19 operating rooms averaged 2.22 +/- 1.13 mG at 30 cm from the main monitor center, 1.29 +/- 0.84 mG at 50 cm, and 1.00 +/- 0.78 mG at the anesthesiologist's standing point. Using 2 mG as a cutoff for high exposure, some operating rooms exceeded this level.",
"effect_direction": "unclear",
"limitations": [
"Health outcomes were not measured (exposure assessment only).",
"Cutoff value (2 mG) was based on prior reports and not justified within the study.",
"Frequency characteristics beyond 'ELF' were not reported.",
"Single-hospital setting may limit generalizability."
],
"evidence_strength": "low",
"confidence": 0.78,
"peer_reviewed_likely": "yes",
"keywords": [
"electromagnetic fields",
"ELF",
"magnetic field density",
"operating room",
"anesthesiologist",
"occupational exposure",
"monitor"
],
"suggested_hubs": [
{
"slug": "occupational-exposure",
"weight": 0.9,
"reason": "Measures EMF exposure levels for anesthesiologists in operating rooms (workplace exposure assessment)."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.3349/ymj.2003.44.1.133
PubMed: 12619186
Related papers
- Fuzzy-Based Bioengineering System for Predicting and Diagnosing Diseases of the Nervous System 2024
- Development of a source-exposure matrix for occupational EMF exposure assessment in the INTEROCC study 2016
- Scoping Review of the Potential Health Effects of Exposure to Extremely Low-Frequency Electric and Magnetic Fields 2019
- Safety Management of Electromagnetic Fields in the Work Environment 2019
- Effect of Occupational Exposure to Low-frequency Electromagnetic Fields on Cataract Development 2025
- Occupational exposure to intermediate frequency and extremely low frequency magnetic fields among personnel working near electronic article surveillance systems 2014