Enhanced occlusion of vessels combining retrievable, detachable coils as differential electrodes with percutaneous, intravascular radiofrequency electrocoagulation. An experimental study.

Investigative radiology · 1996 · Mixed · Low evidence

← Research stream
MixedLow evidencePaper 4212

Complete occlusion was achieved in all cases after a maximum of three consecutive 10-second radiofrequency current applications. Control-side coils remained patent for at least 45 minutes. Gross and histologic examination showed arteries remained intact on bo…

Study snapshot

PublicationInvestigative radiology · 1996
Study typeanimal
Reported directionMixed
Evidence ratingLow

Abstract

RATIONALE AND OBJECTIVES: The authors evaluate the feasibility to accelerate occlusion of high-velocity flow vessels by a combination of transcutaneous coil placement and application of radiofrequency current. METHODS: Piglets (n = 8) were anesthetized and acutely instrumented via cutdowns in both carotid and one brachial arteries. Two identical cylindrically shaped coils (length, 3 mm; outer diameter, 2.4 mm; inner diameter, 1.4 mm) were mounted on titanium-nickel core wire and placed via 3-French Nylon catheters in both iliac arteries. The coils were kept connected to the delivery wire, which is isolated from the surrounding tissue by the catheter. The first-placed system served as control, the contralateral coil was connected to a radiofrequency generator closing electrical circuit via an external indifferent electrode. Angiograms via the brachial artery demonstrated the adequate placement of the coils and the status of the iliac arteries without and with current application. In 6 of the 8 cases, 25 watts of radiofrequency current were applied repeatedly over 10 seconds to the coil on one side at 4-minute intervals until occlusion was demonstrated. In 2 of 8 cases. 25 watts were applied continuously over 30 seconds. The coils were detached from the wire the catheters removed. Additional angiograms were performed after 5, 15, 45, and 60 minutes to show the patency of the control setting. RESULTS: Complete occlusion was achieved in all cases after a maximum of three consecutive applications of current for 10 seconds. The control remained patent for a minimum of 45 minutes. On gross and histologic examination the arteries on both sides remained intact. Disruption and charring occurred only after continuous application of current over 30 seconds. CONCLUSIONS: It is feasible to use detachable coils in conjunction with high-frequency electrocoagulation to promote coil fixation and accelerate occlusion of vessels with high blood flow.

Machine Enhanced Logic

AI evidence extraction

Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.

Study typeanimal
Effect directionmixed
PopulationPiglets
Sample size8
ExposureRepeated 10-second applications at 4-minute intervals until occlusion (max three applications) in 6/8 cases; continuous 30-second application in 2/8 cases
Evidence strengthLowConfidence: 74% · Peer reviewed: yes

Main findings

Complete occlusion was achieved in all cases after a maximum of three consecutive 10-second radiofrequency current applications. Control-side coils remained patent for at least 45 minutes. Gross and histologic examination showed arteries remained intact on both sides, with disruption and charring only after continuous 30-second current application.

Outcomes measured

  • Vessel occlusion (iliac artery)
  • Arterial patency in control side
  • Gross and histologic arterial integrity
  • Tissue disruption/charring with prolonged application

Limitations

  • Experimental animal study (piglets); generalizability to humans not addressed
  • Small sample size (n=8)
  • Radiofrequency frequency not reported
  • Short follow-up (up to 60 minutes)
View raw extracted JSON
{
  "study_type": "animal",
  "exposure": {
    "band": null,
    "source": "intravascular radiofrequency electrocoagulation (radiofrequency generator with coil as electrode)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": "Repeated 10-second applications at 4-minute intervals until occlusion (max three applications) in 6/8 cases; continuous 30-second application in 2/8 cases"
  },
  "population": "Piglets",
  "sample_size": 8,
  "outcomes": [
    "Vessel occlusion (iliac artery)",
    "Arterial patency in control side",
    "Gross and histologic arterial integrity",
    "Tissue disruption/charring with prolonged application"
  ],
  "main_findings": "Complete occlusion was achieved in all cases after a maximum of three consecutive 10-second radiofrequency current applications. Control-side coils remained patent for at least 45 minutes. Gross and histologic examination showed arteries remained intact on both sides, with disruption and charring only after continuous 30-second current application.",
  "effect_direction": "mixed",
  "limitations": [
    "Experimental animal study (piglets); generalizability to humans not addressed",
    "Small sample size (n=8)",
    "Radiofrequency frequency not reported",
    "Short follow-up (up to 60 minutes)"
  ],
  "evidence_strength": "low",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "radiofrequency current",
    "electrocoagulation",
    "intravascular",
    "detachable coils",
    "vessel occlusion",
    "iliac artery",
    "piglet model",
    "angiography",
    "histology"
  ],
  "suggested_hubs": [
    {
      "slug": "engineering",
      "weight": 0.6,
      "reason": "Technique combines detachable coils with radiofrequency electrocoagulation for vascular occlusion; primarily a procedural/technical study rather than environmental EMF exposure."
    }
  ]
}
AI can be wrong. Always verify against the paper.