Improved resolution in the gel electrophoresis of proteins by a periodically interrupted electric field.

Journal of biochemical and biophysical methods · 1996 · Benefit · Low evidence

← Research stream
BenefitLow evidencePaper 4172

In a model protein separation (phycoerythrin vs fluorescein-labeled conalbumin), applying an electric field discontinuously (80 V/cm in 10-s pulses with 120-s pauses) produced about twice the resolution compared with continuous application of 80 V/cm. The sam…

Study snapshot

PublicationJournal of biochemical and biophysical methods · 1996
Study typeengineering
Reported directionBenefit
Evidence ratingLow

Abstract

The capability of the commercial gel electrophoresis apparatus with intermittent scanning of fluorescence (HPGE-1000, LabIntelligence) to provide time-dependent zone dispersion allows one to quantitate resolution. Using a model protein separation, that between phycoerythrin and fluorescein-labeled conalbumin, resolution was compared between separations conducted at a constant field strength of 80 V/cm and one conducted in 10-s pulses of the same field strength, interrupted periodically by 120 s in the absence of an electric field. Resolution was improved by a factor of two in the discontinuous application of the electric field compared to that obtained in its continuous application. Similarly, the intermittent application of 80 V/cm for 10 s, followed by 120-s pauses, gave rise to twice the resolution obtained from a continuous application of 7 V/cm.

Machine Enhanced Logic

AI evidence extraction

Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.

Study typeengineering
Effect directionbenefit
PopulationNot supplied
Sample sizeNot supplied
Exposure10-s pulses interrupted by 120 s without field (also compared to continuous application)
Evidence strengthLowConfidence: 74% · Peer reviewed: yes

Main findings

In a model protein separation (phycoerythrin vs fluorescein-labeled conalbumin), applying an electric field discontinuously (80 V/cm in 10-s pulses with 120-s pauses) produced about twice the resolution compared with continuous application of 80 V/cm. The same intermittent 80 V/cm protocol also yielded about twice the resolution compared with continuous application of 7 V/cm.

Outcomes measured

  • Protein separation resolution (zone dispersion-based quantitated resolution)

Limitations

  • No sample size or number of runs reported in the abstract
  • Study is a methodological/engineering application in gel electrophoresis rather than a health-effects study
  • Limited to a specific model protein separation and specific apparatus/protocol described
View raw extracted JSON
{
  "study_type": "engineering",
  "exposure": {
    "band": null,
    "source": "gel electrophoresis apparatus (electric field)",
    "frequency_mhz": null,
    "sar_wkg": null,
    "duration": "10-s pulses interrupted by 120 s without field (also compared to continuous application)"
  },
  "population": null,
  "sample_size": null,
  "outcomes": [
    "Protein separation resolution (zone dispersion-based quantitated resolution)"
  ],
  "main_findings": "In a model protein separation (phycoerythrin vs fluorescein-labeled conalbumin), applying an electric field discontinuously (80 V/cm in 10-s pulses with 120-s pauses) produced about twice the resolution compared with continuous application of 80 V/cm. The same intermittent 80 V/cm protocol also yielded about twice the resolution compared with continuous application of 7 V/cm.",
  "effect_direction": "benefit",
  "limitations": [
    "No sample size or number of runs reported in the abstract",
    "Study is a methodological/engineering application in gel electrophoresis rather than a health-effects study",
    "Limited to a specific model protein separation and specific apparatus/protocol described"
  ],
  "evidence_strength": "low",
  "confidence": 0.74,
  "peer_reviewed_likely": "yes",
  "keywords": [
    "gel electrophoresis",
    "electric field",
    "pulsed field",
    "intermittent field",
    "protein separation",
    "resolution",
    "zone dispersion",
    "fluorescence scanning"
  ],
  "suggested_hubs": []
}
AI can be wrong. Always verify against the paper.

AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.

Identifiers and source

DOI: 10.1016/s0165-022x(96)00021-8
PubMed: 9029260

Read the original paper →