Spatial and Temporal Variations of a Screening Current Induced Magnetic Field in a Double-Pancake HTS Insert of an LTS/HTS NMR Magnet.
← Research streamExperimental measurements and simulations (critical state model) of the screening current induced magnetic field (SCF) in a Bi2223-tape HTS insert operated at 77 K were compared under self-field, external 5-T background field, and combined excitations. Agreem…
Study snapshot
Abstract
This paper presents experimental and simulation results of a screening current induced magnetic field (SCF) in a high temperature superconductor (HTS) insert that constitutes a low-/high-temperature superconductor (LTS/HTS) NMR magnet. In this experiment, the HTS insert, a stack of 50 double-pancake coils, each wound with Bi2223 tape, was operated at 77 K. A screening current was induced in the HTS insert by three magnetic field sources: 1) a self field from the HTS insert; 2) an external field from a 5-T background magnet; and 3) combinations of 1) and 2). For each field excitation, which induced an SCF, its axial field distribution and temporal variations were measured and compared with simulation results based on the critical state model. Agreement on field profile between experiment and simulation is satisfactory but more work is needed to make the simulation useful for designing shim coils that will cancel the SCF.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
Experimental measurements and simulations (critical state model) of the screening current induced magnetic field (SCF) in a Bi2223-tape HTS insert operated at 77 K were compared under self-field, external 5-T background field, and combined excitations. Agreement between experiment and simulation for the field profile was reported as satisfactory, though additional work was noted as needed to make simulations useful for designing shim coils to cancel SCF.
Outcomes measured
- Axial field distribution of screening current induced magnetic field (SCF)
- Temporal variations of SCF
- Agreement between experimental measurements and critical state model simulations
Limitations
- More work needed to make the simulation useful for designing shim coils that will cancel the SCF
- No health/biological outcomes reported
View raw extracted JSON
{
"study_type": "engineering",
"exposure": {
"band": null,
"source": "NMR magnet (LTS/HTS) / background magnet",
"frequency_mhz": null,
"sar_wkg": null,
"duration": null
},
"population": null,
"sample_size": null,
"outcomes": [
"Axial field distribution of screening current induced magnetic field (SCF)",
"Temporal variations of SCF",
"Agreement between experimental measurements and critical state model simulations"
],
"main_findings": "Experimental measurements and simulations (critical state model) of the screening current induced magnetic field (SCF) in a Bi2223-tape HTS insert operated at 77 K were compared under self-field, external 5-T background field, and combined excitations. Agreement between experiment and simulation for the field profile was reported as satisfactory, though additional work was noted as needed to make simulations useful for designing shim coils to cancel SCF.",
"effect_direction": "unclear",
"limitations": [
"More work needed to make the simulation useful for designing shim coils that will cancel the SCF",
"No health/biological outcomes reported"
],
"evidence_strength": "insufficient",
"confidence": 0.74,
"peer_reviewed_likely": "yes",
"keywords": [
"screening current induced magnetic field",
"SCF",
"high temperature superconductor",
"HTS insert",
"LTS/HTS NMR magnet",
"double-pancake coils",
"Bi2223 tape",
"77 K",
"5-T background magnet",
"critical state model",
"shim coils"
],
"suggested_hubs": []
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1109/tasc.2009.2018102
PubMed: 20401187