Measurement of creatine kinase reaction rate in human brain using magnetization transfer image-selected in vivo spectroscopy (MT-ISIS) and a volume ³¹P/¹H radiofrequency coil in a clinical 3-T MRI system.
← Research streamA new MT 31P MRS method using a volume head 31P coil with ISIS localization was presented for measuring the CK forward reaction rate in a clinical 3-T MRI system with a 30 min acquisition time. The mean k_f measured in 10 healthy volunteers was 0.320 ± 0.075 …
Study snapshot
Abstract
High-energy phosphate metabolism, which allows the synthesis and regeneration of adenosine triphosphate (ATP), is a vital process for neuronal survival and activity. In particular, creatine kinase (CK) serves as an energy reservoir for the rapid buffering of ATP levels. Altered CK enzyme activity, reflecting compromised high-energy phosphate metabolism or mitochondrial dysfunction in the brain, can be assessed using magnetization transfer (MT) MRS. MT (31)P MRS has been used to measure the forward CK reaction rate in animal and human brain, employing a surface radiofrequency coil. However, long acquisition times and excessive radiofrequency irradiation prevent these methods from being used routinely for clinical evaluations. In this article, a new MT (31)P MRS method is presented, which can be practically used to measure the CK forward reaction rate constant in a clinical MRI system employing a volume head (31)P coil for spatial localization, without contamination from the scalp muscle, and an acquisition time of 30 min. Other advantages associated with the method include radiofrequency homogeneity within the regions of interest of the brain using a volume coil with image-selected in vivo spectroscopy localization, and reduction of the specific absorption rate using nonadiabatic radiofrequency pulses for MT saturation. The mean value of k(f) was measured as 0.320 ± 0.075 s(-1) from 10 healthy volunteers with an age range of 18-40 years. These values are consistent with those obtained using earlier methods, and the technique may be used routinely to evaluate energetic processes in the brain on a clinical MRI system.
AI evidence extraction
Structured fields below were generated from the paper’s abstract and metadata to make the evidence easier to inspect. Automated extraction may be incomplete or incorrect; verify important details against the original paper.
Main findings
A new MT 31P MRS method using a volume head 31P coil with ISIS localization was presented for measuring the CK forward reaction rate in a clinical 3-T MRI system with a 30 min acquisition time. The mean k_f measured in 10 healthy volunteers was 0.320 ± 0.075 s^-1, consistent with earlier methods; the method also aimed to reduce specific absorption rate using nonadiabatic RF pulses.
Outcomes measured
- Creatine kinase (CK) forward reaction rate constant (k_f) in human brain measured by MT 31P MRS
View raw extracted JSON
{
"study_type": "exposure_assessment",
"exposure": {
"band": null,
"source": "clinical 3-T MRI system (volume 31P/1H radiofrequency coil)",
"frequency_mhz": null,
"sar_wkg": null,
"duration": "30 min acquisition time"
},
"population": "10 healthy volunteers (age 18–40 years)",
"sample_size": 10,
"outcomes": [
"Creatine kinase (CK) forward reaction rate constant (k_f) in human brain measured by MT 31P MRS"
],
"main_findings": "A new MT 31P MRS method using a volume head 31P coil with ISIS localization was presented for measuring the CK forward reaction rate in a clinical 3-T MRI system with a 30 min acquisition time. The mean k_f measured in 10 healthy volunteers was 0.320 ± 0.075 s^-1, consistent with earlier methods; the method also aimed to reduce specific absorption rate using nonadiabatic RF pulses.",
"effect_direction": "unclear",
"limitations": [],
"evidence_strength": "insufficient",
"confidence": 0.66,
"peer_reviewed_likely": "yes",
"keywords": [
"3-T MRI",
"magnetization transfer",
"31P MRS",
"creatine kinase",
"radiofrequency coil",
"specific absorption rate",
"ISIS localization",
"brain energetics"
],
"suggested_hubs": [
{
"slug": "who-icnirp",
"weight": 0.2,
"reason": "Mentions specific absorption rate (SAR) reduction in an MRI RF method; no explicit policy focus."
}
]
}AI-extracted fields are generated from the abstract and metadata and may be incomplete or incorrect. This material is for informational purposes and is not medical advice.
Identifiers and source
DOI: 10.1002/nbm.1636
PubMed: 21834000
Related papers
- [Effect of Low-Intensity 900 MHz Frequency Electromagnetic Radiation on Rat Brain Enzyme Activities Linked to Energy Metabolism]. 2015
- [Effect of low-intensity 900 MHz frequency electromagnetic radiation on rat liver and blood serum enzyme activities]. 2014
- Pathophysiology of microwave radiation: effect on rat brain. 2012
- A radiofrequency coil to facilitate B₁⁺ shimming and parallel imaging acceleration in three dimensions at 7 T. 2011
- Exposure to Radiofrequency Electromagnetic Fields Enhances Melanin Synthesis by Activating the P53 2024
- Effects of 4G Long-Term Evolution Electromagnetic Fields on Thyroid Hormone Dysfunction and 2024